#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2107D
Results of quality control
Kit: Bordier Affinity Products SA, Fasciola hepatica IgG ELISA, Article
9650 .
Expiry date: 10.11.2022, Control date: 07.03.2022,
Control # 3
No serum blank (average) : 0.114
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.001 |
-0.007 |
0.002 |
0.005 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.069 |
0.083 |
0.072 |
0.069 |
0.073 |
4 |
Weak positive serum
(N° 9650-07) |
0.240 |
0.222 |
0.207 |
0.215 |
0.221 |
11 |
Positive control
serum
(N° 9650-08) |
1.970 |
1.993 |
1.853 |
2.076 |
1.973 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
1.973 |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 9 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.114 |
Yes |
Comments:
Good for use.