Kit: Bordier Affinity Products SA, Toxocara canis IgG ELISA, Article 9200
Expiry date: 20.09.2022, Control date: 19.07.2021,  Control #  2

No serum blank (average) : 0.126

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.002 -0.002 -0.001 0.001 0.000 0
Negative control serum
(N° 9200-06)
0.011 0.007 0.012 0.011 0.010 1
Weak positive serum
(N° 9200-07)
1.061 1.035 1.042 1.040 1.044 56
Positive control serum
(N° 9200-08)
1.852 1.890 1.836 1.938 1.879 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9200-08 > 1.200 1.879 Yes
Absorbance of negative serum 9200-06, as % of positive serum <  8 % 1 % Yes
Absorbance of no serum blank < 0.350 0.126 Yes

Comments:

Good for use.