#9650
Fasciola
hepatica (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2019D
Results of quality control
Kit: Bordier Affinity Products SA, Fasciola hepatica ELISA, Article
9650 .
Expiry date: 30.04.2022, Control date: 02.05.2022,
Control # 4
No serum blank (average) : 0.124
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.000 |
-0.004 |
0.006 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.051 |
0.043 |
0.035 |
0.054 |
0.046 |
2 |
Weak positive serum
(N° 9650-07) |
0.218 |
0.235 |
0.188 |
0.184 |
0.206 |
10 |
Positive control
serum
(N° 9650-08) |
2.032 |
2.154 |
1.887 |
1.921 |
1.998 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
1.998 |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 9 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.124 |
Yes |
Comments:
Good for use.