#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2538D (Expiry date: 02.08.2027)
Production date : 03.12.2025
Results of quality control 1 (after
production), date: 09.12.2025
No serum blank (average) : 0.134
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
-0.003 |
0.000 |
0.006 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.089 |
0.085 |
0.085 |
0.095 |
0.088 |
4 |
Weak positive serum
(N° 9650-07) |
0.379 |
0.373 |
0.386 |
0.390 |
0.382 |
17 |
Positive control
serum
(N° 9650-08) |
2.201 |
2.134 |
2.258 |
2.195 |
2.197 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9650-08 |
> 1.200 |
2.197 |
Yes |
| Absorbance of weak positive serum 9650-07, as % of positive serum |
> 8 % |
17 % |
Yes |
| Absorbance of negative serum 9650-06, as % of positive serum |
< 8 % |
4 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.134 |
Yes |
Comments:
Good for use.