#9750 Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis

Lot 2535R (Expiry date: 26.07.2027)

 

Production date : 27.11.2025

Results of quality control 1 (after production), date: 02.12.2025

No serum blank (average) : 0.131

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 -0.003 0.003 0.000 0.000 0
Negative control serum
(N° 9750-06)
0.050 0.059 0.047 0.060 0.054 2
Weak positive serum
(N° 9750-07)
0.413 0.404 0.423 0.423 0.416 18
Positive control serum
(N° 9750-08)
2.250 2.360 2.224 2.371 2.302 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9750-08 > 1.200 2.302 Yes
Absorbance of weak positive serum 9750-07, as % of positive serum >  10 % 18 % Yes
Absorbance of negative serum 9750-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.131 Yes

Comments:

Good for use.