#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2532S (Expiry date : 18.07.2027)
Production date : 19.11.2025
Results of quality control 1 (after
production), date: 24.11.2025
No serum blank (average) : 0.126
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
-0.005 |
0.008 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.048 |
0.062 |
0.080 |
0.060 |
0.062 |
3 |
Weak positive serum
(N° 9450-07) |
0.898 |
0.845 |
0.881 |
0.914 |
0.884 |
37 |
Positive control serum
(N° 9450-08) |
2.359 |
2.403 |
2.379 |
2.420 |
2.390 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9450-08 |
> 1.200 |
2.390 |
Yes |
| Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
37 % |
Yes |
| Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
3 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.126 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bâtiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|