#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2525A (Expiry date: 07.03.2027)

 

Production date : 08.07.2025

Results of quality control 1 (after 6 months), date: 05.01.2026

No serum blank (average) : 0.132

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 0.002 0.004 -0.003 0.000 0
Negative control serum
(N° 61
00-06)
0.100 0.092 0.128 0.116 0.109 4
Weak positive serum
(N° 61
00-07)
0.732 0.716 0.747 0.731 0.731 29
Positive control serum
(N° 61
00-08)
2.493 2.489 2.477 2.688 2.536 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.536 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 29 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.132 Yes

Comments:

Good for use.