#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2524G (Expiry date: 07.03.2027)
Production date : 08.07.2025
Results of quality control 2 (after
6 months), date: 05.01.2026
No serum blank (average) : 0.132
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.001 |
0.002 |
-0.001 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.124 |
0.117 |
0.135 |
0.113 |
0.122 |
6 |
Weak positive serum
(N° 9350-07) |
1.172 |
1.140 |
1.080 |
1.142 |
1.133 |
55 |
Positive control serum
(N° 9350-08) |
2.100 |
2.119 |
1.997 |
2.100 |
2.079 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9350-08 |
> 1.200 |
2.079 |
Yes |
| Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
55 % |
Yes |
| Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
6 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.132 |
Yes |
Comments:
Good for use.