#9400
Acanthocheilonema
viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis
Lot 2523V (Expiry date: 07.03.2027)
Production date : 08.07.2025
Results of quality control 2 (after
6 months), date: 05.01.2026
No serum blank (average) : 0.133
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.004 |
0.003 |
-0.006 |
0.007 |
0.000 |
0 |
Negative control serum
(N° 9400-06) |
0.061 |
0.067 |
0.074 |
0.067 |
0.067 |
3 |
Weak positive serum
(N° 9400-07) |
0.787 |
0.728 |
0.738 |
0.704 |
0.739 |
37 |
Positive control
serum
(N° 9400-08) |
1.966 |
2.097 |
1.988 |
1.988 |
2.010 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9400-08 |
> 1.200 |
2.010 |
Yes |
| Absorbance of weak positive serum 9400-07, as % of positive serum |
> 19 % |
37 % |
Yes |
| Absorbance of negative serum 9400-06, as % of positive serum |
< 10 % |
3 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.133 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|