#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2519S (Expiry date : 09.02.2027)
Production date : 10.06.2025
Results of quality control 2 (after
6 months), date: 08.12.2025
No serum blank (average) : 0.133
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.009 |
-0.004 |
0.003 |
0.010 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.068 |
0.074 |
0.081 |
0.078 |
0.075 |
4 |
Weak positive serum
(N° 9450-07) |
0.691 |
0.755 |
0.684 |
0.728 |
0.714 |
36 |
Positive control serum
(N° 9450-08) |
2.011 |
2.057 |
1.954 |
2.005 |
2.006 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9450-08 |
> 1.200 |
2.006 |
Yes |
| Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
36 % |
Yes |
| Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
4 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.133 |
Yes |
Comments:
Good for use.