#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2514D (Expiry date: 15.12.2026)
Production date : 16.04.2025
Results of quality control 1 (after
production), date: 23.04.2025
No serum blank (average) : 0.131
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.003 |
-0.002 |
0.001 |
0.004 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.069 |
0.078 |
0.067 |
0.074 |
0.072 |
3 |
Weak positive serum
(N° 9650-07) |
0.315 |
0.320 |
0.293 |
0.320 |
0.312 |
14 |
Positive control
serum
(N° 9650-08) |
2.142 |
2.156 |
2.120 |
2.214 |
2.158 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.158 |
Yes |
Absorbance of weak positive serum 9650-07, as % of positive serum |
> 8 % |
14 % |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 8 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.131 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,