#9650 Fasciola hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 2514D (Expiry date: 15.12.2026)

 

Production date : 16.04.2025

Results of quality control 1 (after production), date: 23.04.2025

No serum blank (average) : 0.131

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 -0.002 0.001 0.004 0.000 0
Negative control serum
(N° 965
0-06)
0.069 0.078 0.067 0.074 0.072 3
Weak positive serum
(N° 965
0-07)
0.315 0.320 0.293 0.320 0.312 14
Positive control serum
(N° 9650
-08)
2.142 2.156 2.120 2.214 2.158 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.158 Yes
Absorbance of weak positive serum 9650-07, as % of positive serum > 8 % 14 % Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.131 Yes

Comments:

Good for use.