Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9310 Echinococcus multilocularis (Em18) IgG ELISA
ELISA kit for human alveolar echinococcosis patient follow-up

Lot 2510F (Expiry date: 19.11.2026)

Production date : 20.03.2025

Results of quality control 2 (after 6 months), date: 16.09.2025

No serum blank (average) : 0.132

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.004 0.006 -0.002 -0.001 0.000 0
Negative control serum
(N° 9310-06)
0.018 0.023 0.011 0.026 0.020 1
Weak positive serum
(N° 9310-07)
0.112 0.109 0.101 0.108 0.108 5
Positive control serum
(N° 9310-08)
2.043 2.233 2.036 2.196 2.127 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9310-08 > 1.200 2.127 Yes
Absorbance of weak positive serum 9310-07, as % of positive serum > 3 % 5 % Yes
Absorbance of negative serum 9310-06, as % of positive serum < 3 % 1 % Yes
Absorbance of no serum blank < 0.350 0.132 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland