#9310
Echinococcus
multilocularis (Em18) IgG ELISA
ELISA kit for human alveolar echinococcosis patient follow-up
Lot 2510F (Expiry date: 19.11.2026)
Production date : 20.03.2025
Results of quality control 1 (after
production), date: 25.03.2025
No serum blank (average) : 0.153
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
0.003 |
-0.003 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9310-06) |
0.025 |
0.023 |
0.030 |
0.029 |
0.027 |
1 |
Weak positive serum
(N° 9310-07) |
0.135 |
0.134 |
0.137 |
0.127 |
0.133 |
6 |
Positive control serum
(N° 9310-08) |
2.262 |
2.394 |
2.406 |
2.380 |
2.361 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9310-08 |
> 1.200 |
2.361 |
Yes |
Absorbance of weak positive serum 9310-07, as % of positive serum
|
> 3 %
|
6 % |
Yes |
Absorbance of negative serum 9310-06, as % of positive serum |
< 3 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.153 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|