Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2509A (Expiry date: 19.10.2026)

 

Production date : 20.02.2025

Results of quality control 1 (after production), date: 25.02.2025

No serum blank (average) : 0.126

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 -0.001 -0.002 0.005 0.000 0
Negative control serum
(N° 61
00-06)
0.048 0.044 0.054 0.051 0.050 2
Weak positive serum
(N° 61
00-07)
0.544 0.517 0.567 0.493 0.531 25
Positive control serum
(N° 61
00-08)
2.175 2.138 2.160 2.105 2.145 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.145 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 25 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.126 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch