#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2508S (Expiry date : 19.10.2026)
Production date : 20.02.2025
Results of quality control 1 (after
production), date: 25.02.2025
No serum blank (average) : 0.132
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.001 |
-0.002 |
-0.003 |
0.004 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.106 |
0.077 |
0.066 |
0.077 |
0.082 |
3 |
Weak positive serum
(N° 9450-07) |
0.789 |
0.810 |
0.806 |
0.807 |
0.803 |
34 |
Positive control serum
(N° 9450-08) |
2.309 |
2.355 |
2.298 |
2.397 |
2.340 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.340 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
34 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.132 |
Yes |
Comments:
Good for use.