Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9700 Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis

Lot 2507C (Expiry date: 21.09.2026)

Production date : 22.01.2025

Results of quality control 1 (after production), date: 28.01.2025

No serum blank (average) : 0.228

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 -0.003 -0.003 0.007 0.000 0
Negative control serum
(N° 9700-06)
0.077 0.074 0.047 0.070 0.067 3
Weak positive serum
(N° 9700-07)
0.443 0.441 0.443 0.421 0.437 17
Positive control serum
(N° 9700-08)
2.501 2.550 2.461 2.544 2.514 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9700-08 > 1.200 2.514 Yes
Absorbance of weak positive serum 9700-07, as % of positive serum > 8 % 17 % Yes
Absorbance of negative serum 9700-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.228 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland