Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9200 Toxocara canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis

Lot 2506T (Expiry date: 21.09.2026)

 

Production date : 22.01.2025

Results of quality control 1 (after production), date: 27.01.2025

No serum blank (average) : 0.128

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 0.003 0.001 -0.001 0.000 0
Negative control serum
(N° 9200-06)
0.043 0.024 0.053 0.043 0.041 2
Weak positive serum
(N° 9200-07)
1.354 1.311 1.365 1.195 1.306 61
Positive control serum
(N° 9200-08)
2.066 2.167 2.104 2.212 2.137 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9200-08 > 1.200 2.137 Yes
Absorbance of weak positive serum 9200-07, as % of positive serum >  32 % 61 % Yes
Absorbance of negative serum 9200-06, as % of positive serum < 6 % 2 % Yes
Absorbance of no serum blank < 0.350 0.128 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch