Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9750 Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis

Lot 2505R (Expiry date: 14.09.2026)

 

Production date : 15.01.2025

Results of quality control 1 (after production), date: 22.01.2025

No serum blank (average) : 0.134

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.004 0.001 0.002 0.001 0.000 0
Negative control serum
(N° 9750-06)
0.027 0.019 0.048 0.139 0.058 3
Weak positive serum
(N° 9750-07)
0.312 0.312 0.442 0.463 0.382 19
Positive control serum
(N° 9750-08)
1.824 1.949 2.161 2.325 2.065 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9750-08 > 1.200 2.065 Yes
Absorbance of weak positive serum 9750-07, as % of positive serum >  10 % 19 % Yes
Absorbance of negative serum 9750-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.134 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch