#9750
Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis
Lot 2505R (Expiry date: 14.09.2026)
Production date : 15.01.2025
Results of quality control 1 (after
production), date: 22.01.2025
No serum blank (average) : 0.134
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.004 |
0.001 |
0.002 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9750-06) |
0.027 |
0.019 |
0.048 |
0.139 |
0.058 |
3 |
Weak positive serum
(N° 9750-07) |
0.312 |
0.312 |
0.442 |
0.463 |
0.382 |
19 |
Positive control serum
(N° 9750-08) |
1.824 |
1.949 |
2.161 |
2.325 |
2.065 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9750-08 |
> 1.200 |
2.065 |
Yes |
Absorbance of weak positive serum 9750-07, as % of positive serum |
> 10 % |
19 % |
Yes |
Absorbance of negative serum 9750-06, as % of positive serum |
< 8 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.134 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +