#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2503G (Expiry date: 07.09.2026)
Production date : 08.01.2025
Results of quality control 4 (after
20 months), date: 09.09.2026
No serum blank (average) : 0.141
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.003 |
0.000 |
0.004 |
0.000 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.089 |
0.088 |
0.103 |
0.086 |
0.091 |
4 |
Weak positive serum
(N° 9350-07) |
1.079 |
1.072 |
1.100 |
1.073 |
1.081 |
50 |
Positive control serum
(N° 9350-08) |
2.082 |
2.208 |
2.197 |
2.168 |
2.164 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9350-08 |
> 1.200 |
2.164 |
Yes |
| Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
50 % |
Yes |
| Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
4 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.141 |
Yes |
Comments:
Good for use.