#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2503G (Expiry date: 07.09.2026)
Production date : 08.01.2025
Results of quality control 3 (after
12 months), date: 30.12.2025
No serum blank (average) : 0.139
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.005 |
-0.001 |
-0.002 |
0.007 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.086 |
0.096 |
0.092 |
0.099 |
0.093 |
4 |
Weak positive serum
(N° 9350-07) |
1.047 |
1.132 |
0.992 |
0.977 |
1.037 |
49 |
Positive control serum
(N° 9350-08) |
2.145 |
2.016 |
2.009 |
2.218 |
2.097 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9350-08 |
> 1.200 |
2.097 |
Yes |
| Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
49 % |
Yes |
| Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
4 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.139 |
Yes |
Comments:
Good for use.