#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2503G (Expiry date: 07.09.2026)
Production date : 08.01.2025
Results of quality control 1 (after
production), date: 13.01.2025
No serum blank (average) : 0.133
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
0.000 |
0.000 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.111 |
0.104 |
0.123 |
0.118 |
0.114 |
5 |
Weak positive serum
(N° 9350-07) |
1.202 |
1.243 |
1.258 |
1.218 |
1.231 |
52 |
Positive control serum
(N° 9350-08) |
2.365 |
2.394 |
2.315 |
2.398 |
2.368 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.368 |
Yes |
Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
52 % |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
5 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.133 |
Yes |
Comments:
Good for use.