#9400 Acanthocheilonema viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis

Lot 2502V (Expiry date: 07.09.2026)

 

Production date : 08.01.2025

Results of quality control 4 (after 20 monhts), date: 09.09.2026

No serum blank (average) : 0.134

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 0.001 0.001 0.000 0.000 0
Negative control serum
(N° 9400-06)
0.072 0.070 0.079 0.065 0.071 3
Weak positive serum
(N° 9400-07)
0.775 0.787 0.801 0.762 0.781 38
Positive control serum
(N° 9400-08)
2.067 2.067 2.055 2.059 2.062 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.062 Yes
Absorbance of weak positive serum 9400-07, as % of positive serum > 19 % 38 % Yes
Absorbance of negative serum 9400-06, as % of positive serum < 10 % 3 % Yes
Absorbance of no serum blank < 0.350 0.134 Yes

Comments:

Good for use.