Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9400 Acanthocheilonema viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis

Lot 2502V (Expiry date: 07.09.2026)

 

Production date : 08.01.2025

Results of quality control 3 (after 12 monhts), date: 30.12.2025

No serum blank (average) : 0.128

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.000 0.001 -0.007 0.007 0.000 0
Negative control serum
(N° 9400-06)
0.081 0.085 0.082 0.095 0.086 4
Weak positive serum
(N° 9400-07)
0.728 0.665 0.687 0.771 0.713 35
Positive control serum
(N° 9400-08)
1.938 2.017 2.077 2.186 2.054 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.054 Yes
Absorbance of weak positive serum 9400-07, as % of positive serum > 19 % 35 % Yes
Absorbance of negative serum 9400-06, as % of positive serum < 10 % 4 % Yes
Absorbance of no serum blank < 0.350 0.128 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland