#9400 Acanthocheilonema viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis

Lot 2502V (Expiry date: 07.09.2026)

 

Production date : 08.01.2025

Results of quality control 1 (after production), date: 13.01.2025

No serum blank (average) : 0.129

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.005 0.004 0.001 0.001 0.000 0
Negative control serum
(N° 9400-06)
0.079 0.096 0.105 0.096 0.094 4
Weak positive serum
(N° 9400-07)
0.848 0.910 0.848 0.895 0.875 41
Positive control serum
(N° 9400-08)
1.973 2.299 2.057 2.192 2.130 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.130 Yes
Absorbance of weak positive serum 9400-07, as % of positive serum > 19 % 41 % Yes
Absorbance of negative serum 9400-06, as % of positive serum < 10 % 4 % Yes
Absorbance of no serum blank < 0.350 0.129 Yes

Comments:

Good for use.