#9550 Entamoeba histolytica
IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis
Lot 2501H (Expiry date: 07.09.2026)
Production date : 08.01.2025
Results of quality control 4 (after
20 months), date: 09.09.2026
No serum blank (average) : 0.144
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.004 |
-0.003 |
0.003 |
0.003 |
0.000 |
0 |
Negative control serum
(N° 9550-06) |
0.092 |
0.089 |
0.098 |
0.087 |
0.092 |
4 |
Weak positive serum
(N° 9550-07) |
0.722 |
0.713 |
0.695 |
0.688 |
0.705 |
33 |
Positive control serum
(N° 9550-08) |
2.196 |
2.214 |
2.125 |
2.105 |
2.160 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9550-08 |
> 1.200 |
2.160 |
Yes |
| Absorbance of weak positive serum 9550-07, as % of positive serum |
> 17 % |
33 % |
Yes |
| Absorbance of negative serum 9550-06, as % of positive serum |
< 12 % |
4 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.144 |
Yes |
Comments:
Good for use.