Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9550 Entamoeba histolytica IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis

Lot 2501H (Expiry date: 07.09.2026)

 

Production date : 08.01.2025

Results of quality control 1 (after production), date: 14.01.2025

No serum blank (average) : 0.141

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.010 0.000 -0.003 0.012 0.000 0
Negative control serum
(N° 9550-06)
0.085 0.094 0.120 0.129 0.107 5
Weak positive serum
(N° 9550-07)
0.804 0.837 0.866 0.904 0.853 37
Positive control serum
(N° 9550-08)
2.160 2.277 2.168 2.515 2.280 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 2.280 Yes
Absorbance of weak positive serum 9550-07, as % of positive serum >  17 % 37 % Yes
Absorbance of negative serum 9550-06, as % of positive serum < 12 % 5 % Yes
Absorbance of no serum blank < 0.350 0.141 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch