#9600
Schistosoma
mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis
Lot 2439B (Expiry date: 12.07.2026)
Production date : 13.11.2024
Results of quality control 1 (after
production), date: 18.11.2024
No serum blank (average) : 0.175
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.015 |
0.009 |
-0.010 |
0.017 |
0.000 |
0 |
Negative control serum
(N° 9600-06) |
0.037 |
0.052 |
0.065 |
0.057 |
0.053 |
2 |
Weak positive serum
(N° 9600-07) |
0.475 |
0.447 |
0.492 |
0.515 |
0.482 |
19 |
Positive control serum
(N° 9600-08) |
2.308 |
2.636 |
2.502 |
2.667 |
2.528 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9600-08 |
> 1.200 |
2.528 |
Yes |
Absorbance of weak positive serum 9600-07, as % of positive serum |
> 10 % |
19 % |
Yes |
Absorbance of negative serum 9600-06, as % of positive serum |
< 8 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.175 |
Yes |
Comments:
Good for use.