Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2439B (Expiry date: 12.07.2026)

 

Production date : 13.11.2024

Results of quality control 1 (after production), date: 18.11.2024

No serum blank (average) : 0.175

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.015 0.009 -0.010 0.017 0.000 0
Negative control serum
(N° 9600-06)
0.037 0.052 0.065 0.057 0.053 2
Weak positive serum
(N° 9600-07)
0.475 0.447 0.492 0.515 0.482 19
Positive control serum
(N° 9600-08)
2.308 2.636 2.502 2.667 2.528 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.528 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 19 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.175 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch