#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2437S (Expiry date : 15.06.2026)
Production date : 16.10.2024
Results of quality control 2 (after
6 months), date: 14.04.2025
No serum blank (average) : 0.133
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.004 |
-0.003 |
0.005 |
-0.006 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.071 |
0.083 |
0.080 |
0.075 |
0.077 |
3 |
Weak positive serum
(N° 9450-07) |
0.727 |
0.770 |
0.757 |
0.758 |
0.753 |
32 |
Positive control serum
(N° 9450-08) |
2.200 |
2.443 |
2.242 |
2.402 |
2.322 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.322 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
32 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.133 |
Yes |
Comments:
Good for use.