#9450 Strongyloïdes ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis

Lot 2437S (Expiry date : 15.06.2026)

Production date : 16.10.2024

Results of quality control 1 (after production), date: 23.10.2024

No serum blank (average) : 0.138

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.007 0.001 -0.002 0.008 0.000 0
Negative control serum
(N° 9450-06)
0.080 0.078 0.088 0.087 0.083 3
Weak positive serum
(N° 9450-07)
0.797 0.848 0.813 0.854 0.828 34
Positive control serum
(N° 9450-08)
2.228 2.659 2.313 2.650 2.462 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.462 Yes
Absorbance of weak positive serum 9450-07, as % of positive serum > 18 % 34 % Yes
Absorbance of negative serum 9450-06, as % of positive serum < 10 % 3 % Yes
Absorbance of no serum blank < 0.350 0.138 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bâtiment Biokema, CH-1023 Crissier, Switzerland