#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2437S (Expiry date : 15.06.2026)
Production date : 16.10.2024
Results of quality control 1 (after
production), date: 23.10.2024
No serum blank (average) : 0.138
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.007 |
0.001 |
-0.002 |
0.008 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.080 |
0.078 |
0.088 |
0.087 |
0.083 |
3 |
Weak positive serum
(N° 9450-07) |
0.797 |
0.848 |
0.813 |
0.854 |
0.828 |
34 |
Positive control serum
(N° 9450-08) |
2.228 |
2.659 |
2.313 |
2.650 |
2.462 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.462 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
34 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.138 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bâtiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|