#9310
Echinococcus
multilocularis (Em18) IgG ELISA
ELISA kit for human alveolar echinococcosis patient follow-up
Lot 2435F (Expiry date: 08.06.2026)
Production date : 09.10.2024
Results of quality control 1 (after
production), date: 17.10.2024
No serum blank (average) : 0.150
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.014 |
0.019 |
-0.007 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9310-06) |
0.036 |
0.032 |
0.044 |
0.031 |
0.035 |
1 |
Weak positive serum
(N° 9310-07) |
0.138 |
0.161 |
0.143 |
0.119 |
0.140 |
6 |
Positive control serum
(N° 9310-08) |
2.361 |
2.617 |
2.260 |
2.449 |
2.421 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9310-08 |
> 1.200 |
2.421 |
Yes |
Absorbance of weak positive serum 9310-07, as % of positive serum
|
> 3 %
|
6 % |
Yes |
Absorbance of negative serum 9310-06, as % of positive serum |
< 3 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.150 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|