#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2434A (Expiry date: 08.06.2026)

 

Production date : 09.10.2024

Results of quality control 1 (after production), date: 22.10.2024

No serum blank (average) : 0.138

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.005 0.000 0.001 0.005 0.000 0
Negative control serum
(N° 61
00-06)
0.088 0.083 0.080 0.082 0.083 4
Weak positive serum
(N° 61
00-07)
0.487 0.467 0.498 0.502 0.488 22
Positive control serum
(N° 61
00-08)
2.216 2.235 2.244 2.239 2.233 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.233 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 22 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.138 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch