Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9300 Echinococcus multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis

Lot 2433M (Expiry date: 08.06.2026)

 

Production date : 09.10.2024

Results of quality control 2 (after 6 months), date: 08.04.2025

No serum blank (average) : 0.131

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.004 0.006 -0.006 0.004 0.000 0
Negative control serum
(N° 9300-06)
0.021 0.028 0.017 0.028 0.023 1
Weak positive serum
(N° 9300-07)
0.590 0.533 0.561 0.613 0.574 25
Positive control serum
(N° 9300-08)
2.107 2.259 2.209 2.430 2.251 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.251 Yes
Absorbance of weak positive serum 9300-07, as % of positive serum > 14 % 25 % Yes
Absorbance of negative serum 9300-06, as % of positive serum < 6 % 1 % Yes
Absorbance of no serum blank < 0.350 0.131 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch