| #9300
Echinococcus
multilocularis (Em2-Em18) IgG ELISA ELISA kit for the diagnosis of human alveolar echinococcosis
 Lot 2433M (Expiry date: 08.06.2026)   Production date : 09.10.2024 Results of quality control 1(after
production), date: 15.10.2024 
 No serum blank (average) : 0.141
 
  
    | Sample | Absorbances of individual wells after blank reduction | Average | % pos. |  
    | No serum blank 
 | -0.005 | 0.003 | 0.001 | 0.002 | 0.000 | 0 |  
    | Negative control serum (N° 9300-06)
 | 0.021 | 0.011 | 0.039 | 0.019 | 0.022 | 1 |  
    | Weak positive serum (N° 9300-07)
 | 0.595 | 0.605 | 0.630 | 0.588 | 0.604 | 27 |  
    | Positive control serum (N° 9300-08)
 | 2.208 | 2.237 | 2.256 | 2.221 | 2.230 | 100 |  Validation criteria: 
  
    |  | Criterion | Measure | Valid |  
    | Absorbance of positive serum 9300-08 | > 1.200 | 2.230 | Yes |  
    | Absorbance of weak positive serum 9300-07, as % of positive serum | > 14 % | 27 % | Yes |  
    | Absorbance of negative serum 9300-06, as % of positive serum | < 6 % | 1 % | Yes |  
    | Absorbance of no serum blank | < 0.350 | 0.141 | Yes |  
Comments:  Good for use.  Bordier Affinity Products SA, Chatanerie 2,
      Bātiment Biokema,
CH-1023 Crissier, SwitzerlandPhone: + 41 21 633 31 67.  Fax: +
      41 21 633 31 78. Email: cb(at)bordier.ch
 |