#9750
Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis
Lot 2432R (Expiry date: 18.05.2026)
Production date : 19.09.2024
Results of quality control 3 (after
12 months), date: 12.09.2025
No serum blank (average) : 0.134
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
0.002 |
-0.003 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9750-06) |
0.047 |
0.050 |
0.044 |
0.052 |
0.048 |
2 |
Weak positive serum
(N° 9750-07) |
0.399 |
0.407 |
0.391 |
0.398 |
0.398 |
18 |
Positive control serum
(N° 9750-08) |
2.026 |
2.315 |
2.054 |
2.281 |
2.169 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9750-08 |
> 1.200 |
2.169 |
Yes |
Absorbance of weak positive serum 9750-07, as % of positive serum |
> 10 % |
18 % |
Yes |
Absorbance of negative serum 9750-06, as % of positive serum |
< 8 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.134 |
Yes |
Comments:
Good for use.