#9750
Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis
Lot 2432R (Expiry date: 18.05.2026)
Production date : 19.09.2024
Results of quality control 2 (after
6 months), date: 17.03.2025
No serum blank (average) : 0.126
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
0.001 |
0.003 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9750-06) |
0.034 |
0.043 |
0.051 |
0.048 |
0.044 |
2 |
Weak positive serum
(N° 9750-07) |
0.447 |
0.423 |
0.405 |
0.402 |
0.419 |
19 |
Positive control serum
(N° 9750-08) |
2.152 |
2.370 |
2.145 |
2.335 |
2.250 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9750-08 |
> 1.200 |
2.250 |
Yes |
Absorbance of weak positive serum 9750-07, as % of positive serum |
> 10 % |
19 % |
Yes |
Absorbance of negative serum 9750-06, as % of positive serum |
< 8 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.126 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|