Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9750 Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis

Lot 2432R (Expiry date: 18.05.2026)

 

Production date : 19.09.2024

Results of quality control 2 (after 6 months), date: 17.03.2025

No serum blank (average) : 0.126

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 0.001 0.003 -0.001 0.000 0
Negative control serum
(N° 9750-06)
0.034 0.043 0.051 0.048 0.044 2
Weak positive serum
(N° 9750-07)
0.447 0.423 0.405 0.402 0.419 19
Positive control serum
(N° 9750-08)
2.152 2.370 2.145 2.335 2.250 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9750-08 > 1.200 2.250 Yes
Absorbance of weak positive serum 9750-07, as % of positive serum >  10 % 19 % Yes
Absorbance of negative serum 9750-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.126 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch