#9750
Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis
Lot 2432R (Expiry date: 18.05.2026)
Production date : 19.09.2024
Results of quality control 1 (after
production), date: 27.09.2024
No serum blank (average) : 0.142
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.009 |
0.008 |
-0.009 |
0.009 |
0.000 |
0 |
Negative control serum
(N° 9750-06) |
0.045 |
0.049 |
0.035 |
0.046 |
0.044 |
2 |
Weak positive serum
(N° 9750-07) |
0.441 |
0.472 |
0.449 |
0.458 |
0.455 |
21 |
Positive control serum
(N° 9750-08) |
2.150 |
2.316 |
2.021 |
2.287 |
2.194 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9750-08 |
> 1.200 |
2.194 |
Yes |
Absorbance of weak positive serum 9750-07, as % of positive serum |
> 10 % |
21 % |
Yes |
Absorbance of negative serum 9750-06, as % of positive serum |
< 8 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.142 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|