Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9750 Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis

Lot 2432R (Expiry date: 18.05.2026)

 

Production date : 19.09.2024

Results of quality control 1 (after production), date: 27.09.2024

No serum blank (average) : 0.142

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.009 0.008 -0.009 0.009 0.000 0
Negative control serum
(N° 9750-06)
0.045 0.049 0.035 0.046 0.044 2
Weak positive serum
(N° 9750-07)
0.441 0.472 0.449 0.458 0.455 21
Positive control serum
(N° 9750-08)
2.150 2.316 2.021 2.287 2.194 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9750-08 > 1.200 2.194 Yes
Absorbance of weak positive serum 9750-07, as % of positive serum >  10 % 21 % Yes
Absorbance of negative serum 9750-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.142 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland