#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2430S (Expiry date : 03.05.2026)
Production date : 04.09.2024
Results of quality control 2 (after
6 months), date: 03.03.2025
No serum blank (average) : 0.132
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.001 |
-0.001 |
0.002 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.077 |
0.073 |
0.084 |
0.083 |
0.079 |
4 |
Weak positive serum
(N° 9450-07) |
0.666 |
0.662 |
0.704 |
0.771 |
0.701 |
32 |
Positive control serum
(N° 9450-08) |
2.062 |
2.135 |
2.147 |
2.350 |
2.173 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.173 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
32 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.132 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bâtiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|