#9650 Fasciola hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 2429D (Expiry date: 13.04.2026)

 

Production date : 14.08.2024

Results of quality control 3 (after 12 months), date: 08.08.2025

No serum blank (average) : 0.130

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.000 0.000 0.000 -0.001 0.000 0
Negative control serum
(N° 965
0-06)
0.067 0.075 0.075 0.073 0.073 3
Weak positive serum
(N° 965
0-07)
0.302 0.307 0.298 0.305 0.303 14
Positive control serum
(N° 9650
-08)
2.095 2.286 2.154 2.219 2.189 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.189 Yes
Absorbance of weak positive serum 9650-07, as % of positive serum > 8 % 14 % Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.130 Yes

Comments:

Good for use.