#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2429D (Expiry date: 13.04.2026)
Production date : 14.08.2024
Results of quality control 3 (after
12 months), date: 08.08.2025
No serum blank (average) : 0.130
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.000 |
0.000 |
0.000 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.067 |
0.075 |
0.075 |
0.073 |
0.073 |
3 |
Weak positive serum
(N° 9650-07) |
0.302 |
0.307 |
0.298 |
0.305 |
0.303 |
14 |
Positive control
serum
(N° 9650-08) |
2.095 |
2.286 |
2.154 |
2.219 |
2.189 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.189 |
Yes |
Absorbance of weak positive serum 9650-07, as % of positive serum |
> 8 % |
14 % |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 8 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.130 |
Yes |
Comments:
Good for use.