#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2429D (Expiry date: 13.04.2026)
Production date : 14.08.2024
Results of quality control 2 (after
6 months), date: 11.02.2025
No serum blank (average) : 0.130
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.003 |
0.003 |
-0.002 |
0.003 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.095 |
0.090 |
0.118 |
0.100 |
0.101 |
4 |
Weak positive serum
(N° 9650-07) |
0.340 |
0.348 |
0.342 |
0.344 |
0.343 |
15 |
Positive control
serum
(N° 9650-08) |
2.227 |
2.411 |
2.202 |
2.292 |
2.283 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.283 |
Yes |
Absorbance of weak positive serum 9650-07, as % of positive serum |
> 8 % |
15 % |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 8 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.130 |
Yes |
Comments:
Good for use.