#9650 Fasciola hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 2429D (Expiry date: 13.04.2026)

 

Production date : 14.08.2024

Results of quality control 2 (after 6 months), date: 11.02.2025

No serum blank (average) : 0.130

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 0.003 -0.002 0.003 0.000 0
Negative control serum
(N° 965
0-06)
0.095 0.090 0.118 0.100 0.101 4
Weak positive serum
(N° 965
0-07)
0.340 0.348 0.342 0.344 0.343 15
Positive control serum
(N° 9650
-08)
2.227 2.411 2.202 2.292 2.283 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.283 Yes
Absorbance of weak positive serum 9650-07, as % of positive serum > 8 % 15 % Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.130 Yes

Comments:

Good for use.