#9650 Fasciola hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 2429D (Expiry date: 13.04.2026)

 

Production date : 14.08.2024

Results of quality control 1 (after production), date: 20.08.2024

No serum blank (average) : 0.121

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 -0.002 0.001 0.001 0.000 0
Negative control serum
(N° 965
0-06)
0.101 0.095 0.087 0.095 0.095 4
Weak positive serum
(N° 965
0-07)
0.357 0.355 0.355 0.328 0.349 16
Positive control serum
(N° 9650
-08)
2.088 2.326 2.168 2.302 2.221 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.221 Yes
Absorbance of weak positive serum 9650-07, as % of positive serum > 8 % 16 % Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.121 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch