#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2428G (Expiry date: 13.04.2026)
Production date : 14.08.2024
Results of quality control 3 (after
12 months), date: 08.08.2025
No serum blank (average) : 0.132
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.005 |
0.000 |
0.002 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.026 |
0.030 |
0.040 |
0.038 |
0.034 |
1 |
Weak positive serum
(N° 9350-07) |
0.986 |
1.066 |
1.044 |
1.044 |
1.035 |
45 |
Positive control serum
(N° 9350-08) |
2.238 |
2.298 |
2.239 |
2.413 |
2.297 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.297 |
Yes |
Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
45 % |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.132 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|