#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2428G (Expiry date: 13.04.2026)
Production date : 14.08.2024
Results of quality control 1 (after
production), date: 19.08.2024
No serum blank (average) : 0.122
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.002 |
-0.004 |
0.002 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.043 |
0.053 |
0.045 |
0.055 |
0.049 |
2 |
Weak positive serum
(N° 9350-07) |
1.135 |
1.126 |
1.139 |
1.223 |
1.155 |
48 |
Positive control serum
(N° 9350-08) |
2.290 |
2.468 |
2.318 |
2.559 |
2.408 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.408 |
Yes |
Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
48 % |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.122 |
Yes |
Comments:
Good for use.