Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9700 Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis

Lot 2427C-R (Expiry date: 21.03.2026)

Production date : 22.07.2024

Results of quality control 2 (after 6 months & rework), date: 17.01.2025

No serum blank (average) : 0.201

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.008 -0.005 0.009 0.004 0.000 0
Negative control serum
(N° 9700-06)
0.090 0.089 0.135 0.126 0.110 5
Weak positive serum
(N° 9700-07)
0.356 0.369 0.424 0.473 0.405 18
Positive control serum
(N° 9700-08)
2.018 2.196 2.182 2.436 2.208 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9700-08 > 1.200 2.208 Yes
Absorbance of weak positive serum 9700-07, as % of positive serum > 8 % 18 % Yes
Absorbance of negative serum 9700-06, as % of positive serum < 8 % 5 % Yes
Absorbance of no serum blank < 0.350 0.201 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland