#9700
Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis
Lot 2427C-R (Expiry date: 21.03.2026)
Production date : 22.07.2024
Results of quality control 1 (after
production & rework), date: 11.09.2024
No serum blank (average) : 0.201
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.007 |
-0.003 |
-0.004 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9700-06) |
0.110 |
0.105 |
0.097 |
0.076 |
0.097 |
4 |
Weak positive serum
(N° 9700-07) |
0.450 |
0.447 |
0.414 |
0.417 |
0.432 |
19 |
Positive control
serum
(N° 9700-08) |
2.245 |
2.531 |
2.197 |
2.348 |
2.331 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9700-08 |
> 1.200 |
2.331 |
Yes |
Absorbance of weak positive serum 9700-07, as % of positive serum
|
> 8 % |
19 % |
Yes
|
Absorbance of negative serum 9700-06, as % of positive serum |
< 8 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.201 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|