#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2426T-R (Expiry date: 21.03.2026)
Production date : 22.07.2024
Results of quality control 1 (after
production & rework), date: 10.09.2024
No serum blank (average) : 0.133
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
-0.001 |
0.001 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.043 |
0.046 |
0.050 |
0.039 |
0.044 |
2 |
Weak positive serum
(N° 9200-07) |
1.471 |
1.489 |
1.491 |
1.564 |
1.504 |
66 |
Positive control serum
(N° 9200-08) |
2.243 |
2.322 |
2.240 |
2.283 |
2.272 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9200-08 |
> 1.200 |
2.272 |
Yes |
Absorbance of weak positive serum 9200-07, as % of positive serum |
> 32 % |
66 % |
Yes |
Absorbance of negative serum 9200-06, as % of positive serum |
< 6 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.133 |
Yes |
Comments:
Good for use.