#9400 Acanthocheilonema viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis

Lot 2424V (Expiry date: 07.03.2026)

 

Production date : 08.07.2024

Results of quality control 2 (after 6 months), date: 03.01.2025

No serum blank (average) : 0.166

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.011 0.037 -0.029 0.003 0.000 0
Negative control serum
(N° 9400-06)
0.101 0.092 0.060 0.079 0.083 3
Weak positive serum
(N° 9400-07)
0.873 0.924 0.826 0.841 0.866 36
Positive control serum
(N° 9400-08)
2.221 2.568 2.191 2.555 2.384 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.384 Yes
Absorbance of weak positive serum 9400-07, as % of positive serum > 19 % 36 % Yes
Absorbance of negative serum 9400-06, as % of positive serum < 10 % 3 % Yes
Absorbance of no serum blank < 0.350 0.166 Yes

Comments:

Good for use.