#9400 Acanthocheilonema viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis

Lot 2424V (Expiry date: 07.03.2026)

 

Production date : 08.07.2024

Results of quality control 1 (after production), date: 10.07.2024

No serum blank (average) : 0.123

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 0.001 0.001 0.000 0.000 0
Negative control serum
(N° 9400-06)
0.087 0.106 0.091 0.092 0.094 4
Weak positive serum
(N° 9400-07)
0.884 0.822 0.873 0.880 0.865 39
Positive control serum
(N° 9400-08)
2.183 2.243 2.138 2.363 2.232 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.232 Yes
Absorbance of weak positive serum 9400-07, as % of positive serum > 19 % 39 % Yes
Absorbance of negative serum 9400-06, as % of positive serum < 10 % 4 % Yes
Absorbance of no serum blank < 0.350 0.123 Yes

Comments:

Good for use.