#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2423S-R (Expiry date : 02.03.2026)
Production date : 03.07.2024
Results of quality control 3 (after
12 months & rework), date: 27.06.2025
No serum blank (average) : 0.135
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.002 |
0.003 |
-0.002 |
-0.003 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.076 |
0.082 |
0.089 |
0.094 |
0.085 |
4 |
Weak positive serum
(N° 9450-07) |
0.750 |
0.778 |
0.728 |
0.763 |
0.755 |
32 |
Positive control serum
(N° 9450-08) |
2.319 |
2.460 |
2.328 |
2.469 |
2.394 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.394 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
32 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.135 |
Yes |
Comments:
Good for use.