#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2423S-R (Expiry date : 02.03.2026)
Production date : 03.07.2024
Results of quality control 1 (after
production & rework), date: 03.09.2024
No serum blank (average) : 0.142
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.002 |
0.000 |
0.001 |
-0.003 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.132 |
0.126 |
0.124 |
0.118 |
0.125 |
5 |
Weak positive serum
(N° 9450-07) |
0.900 |
0.934 |
0.863 |
0.942 |
0.910 |
34 |
Positive control serum
(N° 9450-08) |
2.548 |
2.900 |
2.527 |
2.813 |
2.697 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.697 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
34 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
5 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.142 |
Yes |
Comments:
Good for use.