#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2422B-R (Expiry date: 25.02.2026)

 

Production date : 26.06.2024

Results of quality control 3 (after 12 months & rework), date: 23.06.2025

No serum blank (average) : 0.144

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.004 0.005 -0.005 0.005 0.000 0
Negative control serum
(N° 9600-06)
0.050 0.067 0.065 0.075 0.064 3
Weak positive serum
(N° 9600-07)
0.462 0.468 0.447 0.489 0.466 19
Positive control serum
(N° 9600-08)
2.392 2.425 2.260 2.535 2.403 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.403 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 19 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.144 Yes

Comments:

Good for use.